국산 자동화 장비(Gamma Pro)의 결과보고시간 단축에 대한 유용성 평가

이용수 0

영문명
Evaluation of Reliability about Short TAT (Turn-Around Time) of Domestic Automation Equipment (Gamma Pro)
발행기관
대한핵의학기술학회
저자명
오윤정(Yun Jeong Oh) 김지영(Ji Young Kim) 석재동(Jae Dong Seok)
간행물 정보
『핵의학기술』Vol.14 No.2, 197~202쪽, 전체 6쪽
주제분류
의약학 > 방사선과학
파일형태
PDF
발행일자
2010.10.30
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose: Recently, many hospitals have been tried to increase the satisfaction of the outpatients through blood-gathering, exam, result notice and process in a day. Each laboratory has been used the automatic equipment for the rapid requests of the result notice and the increase of the reliability and efficiency. Current automatic equipments that have been limited short TAT(Turn-Around Time)because of the restricted batch lists and 1 tip-5 detectors. The Gamma Pro which is made in Korea to improve the shortcomings of existing automation equipment, complemented with capacity to perform a wide range of domestic automation equipment. In this study, we evaluated the usefulness and reliability of short TAT by comparing Gamma Pro with current automatic equipment. Materials and Methods: We studied the correlation between Gamma Pro and RIA-mat 280 using the respective 100 specimens of low or high density to the patients who were requested the thyroid hormone test (Total T3, TSH and Free T4) in Samsung Medical Center Sep. 2009. To evaluate the split-level Gamma Pro, First, we measured accuracy and carry over on the tips. Second, the condition of optimal incubation was measured by the RPM (Revolution Per Minute) and revolution axis diameter on the incubator. For the analysis for the speed of the specimen-processing, TAT was investigated with the results in a certain time. Result: The correlation coefficients (R2) between the Gamma Pro and RIA-mat 280 showed a good correlation as T3 (0.98), TSH (0.99), FT4 (0.92). The coefficient of variation (C.V) and accuracy was 0.38 % and 98.3 % at tip 1 and 0.39 % and 98.6 % at tip 2. Carry over showed 0.80 % and 1.04% at tip 1 and tip 2, respectively. These results indicate that tips had no effect on carry over contamination. At the incubator condition, we found that the optimal condition was 1.0mm of diameter at 600RPM in 1.0mm and 1.5mm of at 500RPM or 1.0mm and 1.5 mm of diameter at 600 RPM. the Gamma Pro showed that the number of exam times were increased as maximum 20 times/day comparing to 6 times/day by current automatic equipment. These results also led to the short TAT from 4.20 hour to 2.19 hours in whole processing. Conclusion: The correlation of between the Gamma Pro and RIA-mat 280 was good and has not carry over contamination in tips. The domestic automation equipment (Gamma Pro) decreases the TAT in whole test comparing to RIA-280. These results demonstrate that Gamma Pro has a good efficiency, reliability and practical usefulness, which may contribute to the excellent skill to process the large scale specimens.

목차

서론
측정방법 및 대상
결과
고찰 및 결론
요약
참고문헌

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

오윤정(Yun Jeong Oh),김지영(Ji Young Kim),석재동(Jae Dong Seok). (2010).국산 자동화 장비(Gamma Pro)의 결과보고시간 단축에 대한 유용성 평가. 핵의학기술, 14 (2), 197-202

MLA

오윤정(Yun Jeong Oh),김지영(Ji Young Kim),석재동(Jae Dong Seok). "국산 자동화 장비(Gamma Pro)의 결과보고시간 단축에 대한 유용성 평가." 핵의학기술, 14.2(2010): 197-202

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제