두개부와 흉부 CT 검사시 Bismuth 차폐재를 이용한 안구의 렌즈와 유방의 선량감소에 대한 평가

이용수 38

영문명
The analysis for the reduction of radiation dose of breast and lens in eyeballs during the CT examination of cranium and chest using Bismuth shields
발행기관
대한CT영상기술학회
저자명
변정인(Jung in Byun) 남윤철(Yoon chul Nam) 이홍(Hong Lee) 신상보(Sang bo Shin) 한동균(Dong kyoon Han)
간행물 정보
『대한CT영상기술학회지』대한전산화단층기술학회지 제10권 제1호, 195~201쪽, 전체 7쪽
주제분류
의약학 > 방사선과학
파일형태
PDF
발행일자
2008.04.30
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose Bismuth shield can filter twice in that penetrates X-ray, which is essential in diagnosis, and it also partially decreases skin dose. Hence, it is necessary to study the effects in reduction of absorbed dose on highly sensitive organs, which are included in the regions of cranial and chest CT exams. Materials and methods All exams were proceeded on LlghtSpeed 16MDCT (GE Medical System, Milwaukee, U.S.A.), Rando phantom (Model RAN-110, Churchin associate LID., USA) and Glass dosimetry system(GD-351,8.5mm, FGD-1000) for the measurement of radiation dose. In addition, AttenuRad radiation protection devices of eyeballs & breast (F&L Medical Products Co.) are used for protection of sensitive organs. In cranial examination, the brain CT protocols was used for the examinations, and the summary of it is as followed: The glass radiation elements were put on left and right lens of phantom eyeballs, and the radiation before and after the use of Bismuth shield were measured twice when 120kVp of current were changed into 150mAs, 200mAs, and 200mAs, and when 250mAs were changed into 100kVp, 120kVp, and 140kVp. In chest examination, low-dose protocol and chest routine protocol are used for the examination. The conditions of low-dose protocol are as followed:120kVp, noise index 11.57(10~50mA), tube rotation time 0.8sec, pitch 1.375, and table speed 13.75mm/sec. Also, the conditions of chest routine CT protocol are 120kVp, noise index 15.86(160~250mA), tube rotation time 0.8sec, pitch 1.35, and table speed 13.5mm/sec. The glass elements were put on the both nipples of phantom and the radiation before and after the use of bismuth shield were measured twice. When the shield is used, rubber-made spacer(1cm) should be on the breasts, followed by bismuth shield, because it decreases not only the scattering ray absorption into breast wall caused by bismuth shield but image artifact potentially. Results The result of the measurement of orbit shielding during cranial examination. The results of measurements of the radiation(from 120kVp to 150, 200, 250 and 300mAs) before and after the use of bismuth shield are as followed: from 18.4±0.3mGy to 10.25±0.5rnGy(44.3% decrease) at 150mAs; from 23.9±0.9mGy to 13.9±0.1mGy(41.5% decrease) at 200mAs; from 37.3±0.3mGy to 19.8±0.9mGy(47.1% decrease) at 250mAs; and from 47.8±3.4mGy to 24.4±1.4mGy(46.8% decrease) at 300mAs. On the other hand, the measurements of radiation(from 250mAs to 100, 120, and 140 kVp) before and after the use of bismuth shield are as followed: from 22.9±0.2mGy to 22.9±0.2mGy(51.2% decrease) at 100kVp; from 34.4±0.9mGy to 19.5±0.2mGy(43.3% decrease) at 120kVp; and from 49.5 ±4.4mGy to 29±1.2mGy(41.4% decrease) at 140kVp. The result of the measurement of orbit shielding during breast examination. In low-dose protocol, the results of measurements of the radiation before and after the use of bismuth shield are 3.16±0.02mGy and 2.44±0.03mGy(22.78% decrease), respectively, and in chest routine protocol, the results were 14.45±1.08mGy and 11.93±0.21mGy(17.43% decrease), respectively. Conclusion In cranial examination, there was not much change in the radiation dose of eye lens with an application of bismuth shield, even increases of mAs with constant kVp. On the other hand, the rate of reduction of radiation dose by bismuth shield is seen with constant mAs and increase of kVp. Due to the use of bismuth shield, radiation dose has been decreased about 22.8%in low-dose study, and 17.4% in chest routine study. In our hospital, the use of bismuth shield of orbit is not applied to adults, but applied to some infants only when brain CT exam was operated to them. The breast shield is applied, as a test, to women under 50 years old. According to this study, the results show obviously that the use of the bismuth shield can reduce radiation dose of highly radiation-sensitive organs. As mentioned before, the amount of radiation that eye balls received at 250mAs(at

목차

Abstract
I. 서론
II. 재료 및 방법
III. 결과
IV. 고찰
V. 결론
참고문헌

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

변정인(Jung in Byun),남윤철(Yoon chul Nam),이홍(Hong Lee),신상보(Sang bo Shin),한동균(Dong kyoon Han). (2008).두개부와 흉부 CT 검사시 Bismuth 차폐재를 이용한 안구의 렌즈와 유방의 선량감소에 대한 평가. 대한CT영상기술학회지, 10 (1), 195-201

MLA

변정인(Jung in Byun),남윤철(Yoon chul Nam),이홍(Hong Lee),신상보(Sang bo Shin),한동균(Dong kyoon Han). "두개부와 흉부 CT 검사시 Bismuth 차폐재를 이용한 안구의 렌즈와 유방의 선량감소에 대한 평가." 대한CT영상기술학회지, 10.1(2008): 195-201

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제