학술논문
나선형 CT와 16 Slice MDCT의 Brain과 Abdome 검사 시 산란선에 관한 연구
이용수 48
- 영문명
- Study on Scattered Ray in Spiral CT and 16-Slice MDCT of Brain and Abdomen
- 발행기관
- 대한CT영상기술학회
- 저자명
- 윤재혁(Jae hyeok Yun) 류병규(Byeong gyu Ryu) 이관원(Gwang won Lee) 김동수(Dong su Kim) 장양선(Yang seon Jang) 구양수(Yang su Ku)
- 간행물 정보
- 『대한CT영상기술학회지』대한전산화단층기술학회지 제10권 제1호, 203~210쪽, 전체 8쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2008.04.30
4,000원
구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.
국문 초록
영문 초록
Purpose
The present study purposed to measure the degree of scattered ray of radiation in spiral CT and 16-Slice MDCT of brain and abdomen using Rando phantom and to examine the change in non-patients’ exposure to scattered ray according to distance between the central beam and the patient and the reduction of exposure according to the thickness of a protection outfit for shielding from radiation.
Materials and methods
We measured the change of scattered ray on non-patients according to distance between the brain beam of CT and the patient and the change of scattered ray according to the thickness of an apron for shielding from scattered ray, assuming a critically ill patient and an infant on Somatom Plus 4 and MDCT table at the present hospital using Rando phantom under the same setting of cerebral and abdominal CT scan parameters. In order to reduce errors in measurements, each factor was measured three times and the mean value was obtained. The height of the survey meter probe was 111.5cm, which is ordinary adults’ iliac crest level.
Results
A. In the result of brain test with Plus 4, the reduction rate of scattered ray at a distance of 50cm according to distance from the central beam without an apron (15.5mR) and with an apron on:
1. When Pb equivalent was 2.5mm, the reduction rate of scattered ray was 91.55% at 50cm, 94.82% at 100cm, 98.64% at 150cm, and 99.38% at 200cm.
2. When Pb equivalent was 5.0mm, the reduction rate of scattered ray was 98.50% at 50cm, 99.26% at 100cm, 99.58% at 150cm, and 99.70% at 200cm.
3. When Pb equivalent was 7.5mm, the reduction rate of scattered ray was 98.87% at 50cm, 99.37% at 100cm, 99.79% at 150cm, and 99.80% at 200cm.
B. In the result of brain test with 16-Slice MDCT, the reduction rate of scattered ray at a distance of 50cm according to distance from the central beam without an apron (16.47mR) and with an apron on:
1. When Pb equivalent was 2.5mm, the reduction rate of scattered ray was 95.40% at 50cm, 97.69% at 100cm, 98.84% at 150cm, and 99.33% at 200cm.
2. When Pb equivalent was 5.0mm, the reduction rate of scattered ray was 98.54% at 50cm, 99.18% at 100cm, 98.54% at 150cm, and 99.66% at 200cm.
3. When Pb equivalent was 7.5mm, the reduction rate of scattered ray was 98.10% at 50cm, 99.55% at 100cm, 99.86% at 150cm, and 99.78% at 200cm.
C. In the result of abdomen test with Plus 4, the reduction rate of scattered ray at a distance of 50cm according to distance from the central beam without an apron (61.47mR) and with an apron on:
1. When Pb equivalent was 2.5mm, the reduction rate of scattered ray was 91.52% at 50cm, 97.15% at 100cm, 98.74% at 150cm, and 99.40% at 200cm.
2. When Pb equivalent was 5.0mm, the reduction rate of scattered ray was 98.48% at 50cm, 99.19% at 100cm, 99.60% at 150cm, and 99.72% at 200cm.
3. When Pb equivalent was 7.5mm, the reduction rate of scattered ray was 98.81% at 50cm, 99.49% at 100cm, 99.71% at 150cm, and 99.82% at 200cm.
D. In the result of abdomen test with 16-Slice MDCT, the reduction rate of scattered ray at a distance of 50cm according to distance from the central beam without an apron (79.47mR) and with an apron on:
1. When Pb equivalent was 2.5mm, the reduction rate of scattered ray was 96.56% at 50cm, 98.64% at 100cm, 99.34% at 150cm, and 99.69% at 200cm.
2. When Pb equivalent was 5.0mm, the reduction rate of scattered ray was 98.96% at 50cm, 99.45% at 100cm, 99.72% at 150cm, and 99.85% at 200cm.
3. When Pb equivalent was 7.5mm, the reduction rate of scattered ray was 99.56% at 50cm, 99.75% at 100cm, 99.86% at 150cm, and 99.94% at 200cm.
Conclusion
According to the results of this study, scattered ray decreased with the increase of distance. The scattered ray shielding effect was highest when the Pb equivalent of the radiation shielding outfit was 7.5mm. However, the reduction rate of scattered ray with a radiation shielding outfit at thickness of 5.0mm was higher (+4% on the average) than that with an outfit a
목차
Abstract
I. 목적
II. 대상 및 방법
III. 결과
IV. 고찰
V. 결론
참고문헌
해당간행물 수록 논문
- CT Mammography검사 시 자세변화에 따른 영상평가 및 적정선량에 관한 연구
- MDCT에 사용되는 조영제의 부작용 감소방안에 대한 고찰
- Effective mAs 적용 CT에서 Pitch Factor 변화에 따른 Image Quality의 비교
- 조영제 부작용 환자 등록 전산프로그램 및 조영제 부작용 표준진료지침개발 후 부작용사례분석
- MDCT를 이용한 복부 3-phase검사시 추적검사환자의 피폭선량 감소를 위한 Bolus triggering에 대한 고찰
- 검진센터를 방문한 성인에서 64-MDCT를 이용한 뇌동맥 관상동맥 질환의 발견율에 대한 고찰
- DS(Dual Source) CT를 이용한 심장검사 시 Pitch선택에 따른 영향 분석
- Dual Source CT에서 Coronary검사의 방사선량 비교분석
- Low Dose CT Colonography(LDCTC)를 이용한 Automatic Exposure Controll(AEC)와 Beam Collimation의 변화에 따른 유용성 평가와 Effective Dose의 평가치에 대한 연구
- MDCT을 이용한 Cardica CT 검사에서 ECG-pulsing기능의 사용 경험 및 고찰
- Cardiac CT에서 Dual injector의 Dual flow technique을 이용한 Contrast media의 Optimize admix rate에 관한 연구
- Dual flow 기법을 이용한 Triphasic injection의 유용성
- MDCT를 이용한 Head Angiography검사 시 Contrast Volume 감소에 관한 연구
- MDCT를 이용한 Neck Angiography검사 시 조영제량 감소에 따른 화질 향상에 관한 연구
- 상행대동맥의 조영증강시간에 영향을 주는 인자 관한 연구
- Coronary CT Angiography에서 Cardio ImageXact
- Low dose CT Virtual Colonography의 유용성 평가
- MDCT의 다양한 알고리즘과 관상동맥내 삽입된 스텐트 내경측정의 영향
- 나선형 CT와 16 Slice MDCT의 Brain과 Abdome 검사 시 산란선에 관한 연구
- 급성 장출혈 환자에서 MDCT의 유용성
- Cardiac CTA시 진단영역 확대를 위한 Heart rate에 따른 조영제 주입방법에 관한 연구
- CT 3차원 영상의 의료 수가의 추가적인 분리 적용의 필요성에 대한 제언
- CT 유도하 대동맥 주변 임파절 흡입과 생검에 대한 고찰
- 저선량 CT Guided Interventional Study의 유용성 평가
- Care Dose 4D를 이용한 흉부검사의 노출조건의 적정성 연구
- 전산화 단층촬영에 사용되는 조영제의 부작용 발생에 대한 분석
- 64VCT를 이용한 관상동맥 검사시 z-axis Automatic CT tube current modulation을 통한 방사선량 감소방안에 대한 연구
- 두개부와 흉부 CT 검사시 Bismuth 차폐재를 이용한 안구의 렌즈와 유방의 선량감소에 대한 평가
- Spiral CT에서 Interpolation이 선량증가에 미치는 영향에 관한 연구
- CT 검사 시 촬영조건에 따른 Image Quality와 Dose에 관한 연구
- Coronary Angio CT에서 심박동수에 의한 조영증강의 추이
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 의료용 인공지능에 대한 동향 분석 및 방사선사 인식조사
- CT contrast 검사 시 중심 정맥관을 이용한 조영제 사용에 대한 연구
- CT조영제 부작용 고위험군에 대한 전처치의 유용성
의약학 > 방사선과학분야 NEW
- 초음파 조직검사에 사용되는 Biopsy Gun Needle의 재질에 따른 반사율 연구
- 방사선에 대한 지식 및 인식도 연구
- 입원환자 일반촬영 이용량 및 피폭선량: 2018년 입원환자데이터
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!