나선형 CT와 16 Slice MDCT의 Brain과 Abdome 검사 시 산란선에 관한 연구

이용수 48

영문명
Study on Scattered Ray in Spiral CT and 16-Slice MDCT of Brain and Abdomen
발행기관
대한CT영상기술학회
저자명
윤재혁(Jae hyeok Yun) 류병규(Byeong gyu Ryu) 이관원(Gwang won Lee) 김동수(Dong su Kim) 장양선(Yang seon Jang) 구양수(Yang su Ku)
간행물 정보
『대한CT영상기술학회지』대한전산화단층기술학회지 제10권 제1호, 203~210쪽, 전체 8쪽
주제분류
의약학 > 방사선과학
파일형태
PDF
발행일자
2008.04.30
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose The present study purposed to measure the degree of scattered ray of radiation in spiral CT and 16-Slice MDCT of brain and abdomen using Rando phantom and to examine the change in non-patients’ exposure to scattered ray according to distance between the central beam and the patient and the reduction of exposure according to the thickness of a protection outfit for shielding from radiation. Materials and methods We measured the change of scattered ray on non-patients according to distance between the brain beam of CT and the patient and the change of scattered ray according to the thickness of an apron for shielding from scattered ray, assuming a critically ill patient and an infant on Somatom Plus 4 and MDCT table at the present hospital using Rando phantom under the same setting of cerebral and abdominal CT scan parameters. In order to reduce errors in measurements, each factor was measured three times and the mean value was obtained. The height of the survey meter probe was 111.5cm, which is ordinary adults’ iliac crest level. Results A. In the result of brain test with Plus 4, the reduction rate of scattered ray at a distance of 50cm according to distance from the central beam without an apron (15.5mR) and with an apron on: 1. When Pb equivalent was 2.5mm, the reduction rate of scattered ray was 91.55% at 50cm, 94.82% at 100cm, 98.64% at 150cm, and 99.38% at 200cm. 2. When Pb equivalent was 5.0mm, the reduction rate of scattered ray was 98.50% at 50cm, 99.26% at 100cm, 99.58% at 150cm, and 99.70% at 200cm. 3. When Pb equivalent was 7.5mm, the reduction rate of scattered ray was 98.87% at 50cm, 99.37% at 100cm, 99.79% at 150cm, and 99.80% at 200cm. B. In the result of brain test with 16-Slice MDCT, the reduction rate of scattered ray at a distance of 50cm according to distance from the central beam without an apron (16.47mR) and with an apron on: 1. When Pb equivalent was 2.5mm, the reduction rate of scattered ray was 95.40% at 50cm, 97.69% at 100cm, 98.84% at 150cm, and 99.33% at 200cm. 2. When Pb equivalent was 5.0mm, the reduction rate of scattered ray was 98.54% at 50cm, 99.18% at 100cm, 98.54% at 150cm, and 99.66% at 200cm. 3. When Pb equivalent was 7.5mm, the reduction rate of scattered ray was 98.10% at 50cm, 99.55% at 100cm, 99.86% at 150cm, and 99.78% at 200cm. C. In the result of abdomen test with Plus 4, the reduction rate of scattered ray at a distance of 50cm according to distance from the central beam without an apron (61.47mR) and with an apron on: 1. When Pb equivalent was 2.5mm, the reduction rate of scattered ray was 91.52% at 50cm, 97.15% at 100cm, 98.74% at 150cm, and 99.40% at 200cm. 2. When Pb equivalent was 5.0mm, the reduction rate of scattered ray was 98.48% at 50cm, 99.19% at 100cm, 99.60% at 150cm, and 99.72% at 200cm. 3. When Pb equivalent was 7.5mm, the reduction rate of scattered ray was 98.81% at 50cm, 99.49% at 100cm, 99.71% at 150cm, and 99.82% at 200cm. D. In the result of abdomen test with 16-Slice MDCT, the reduction rate of scattered ray at a distance of 50cm according to distance from the central beam without an apron (79.47mR) and with an apron on: 1. When Pb equivalent was 2.5mm, the reduction rate of scattered ray was 96.56% at 50cm, 98.64% at 100cm, 99.34% at 150cm, and 99.69% at 200cm. 2. When Pb equivalent was 5.0mm, the reduction rate of scattered ray was 98.96% at 50cm, 99.45% at 100cm, 99.72% at 150cm, and 99.85% at 200cm. 3. When Pb equivalent was 7.5mm, the reduction rate of scattered ray was 99.56% at 50cm, 99.75% at 100cm, 99.86% at 150cm, and 99.94% at 200cm. Conclusion According to the results of this study, scattered ray decreased with the increase of distance. The scattered ray shielding effect was highest when the Pb equivalent of the radiation shielding outfit was 7.5mm. However, the reduction rate of scattered ray with a radiation shielding outfit at thickness of 5.0mm was higher (+4% on the average) than that with an outfit a

목차

Abstract
I. 목적
II. 대상 및 방법
III. 결과
IV. 고찰
V. 결론
참고문헌

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

윤재혁(Jae hyeok Yun),류병규(Byeong gyu Ryu),이관원(Gwang won Lee),김동수(Dong su Kim),장양선(Yang seon Jang),구양수(Yang su Ku). (2008).나선형 CT와 16 Slice MDCT의 Brain과 Abdome 검사 시 산란선에 관한 연구. 대한CT영상기술학회지, 10 (1), 203-210

MLA

윤재혁(Jae hyeok Yun),류병규(Byeong gyu Ryu),이관원(Gwang won Lee),김동수(Dong su Kim),장양선(Yang seon Jang),구양수(Yang su Ku). "나선형 CT와 16 Slice MDCT의 Brain과 Abdome 검사 시 산란선에 관한 연구." 대한CT영상기술학회지, 10.1(2008): 203-210

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제