혈중 Estradiol 농도 측정을 위한 여러 방사면역측정 검사키트 및 화학면역발광 검사법의 성능평가 및 상호비교

이용수 0

영문명
Quality Assessment and Comparison of Several Radioimmunoassay Kits and Chemiluminescence Immunoassay Methods for Evaluating Serum Estradiol
발행기관
대한핵의학기술학회
저자명
최성희(Sung Hee Choi) 노경운(Gyeong Woon Noh) 김진의(Jin Eui Kim) 송요성(Yoo Sung Song) 팽진철(Jin Chul Paeng) 강건욱(Keon Wook Kang) 이동수(Dong Soo Lee)
간행물 정보
『핵의학기술』Vol.19 No.1, 72~80쪽, 전체 9쪽
주제분류
의약학 > 방사선과학
파일형태
PDF
발행일자
2015.05.30
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose Serum estradiol (E₂) measurement is requested for diagnosing menstrual cycles, ovulation induction, infertility, and menopause. E₂ is measured using several methods and kits including radioimmunoassay (RIA) and chemiluminescece immunoassay (CLIA). The purpose of this study was to evaluate quality of these methods and to compare them with each other. Materials and Methods Seven radioimmunoassay kits and two CLIA methods were included in the analysis. Using standard samples and patient samples, intra-assay precision, inter-assay precision, correlation between other methods, sensitivity, and recovery rate were evaluated. Results For all tested kits and methods, coefficients of variance (CVs) of intra-assay precision test were 10.9~13.6% in low-level samples and less than 10% in medium and high-level samples. CVs of inter-assay precision test were 10.8~12.3% in low-level samples and less than 10% in medium and high-level samples with all tested kits and methods. Recovery rates were 92.7±12.4% for SIEMENS, 101.4±18.4% for DIAsource, 95.1±11.5% for AMP, 108.4±18.5% for BECKMAN COULTER, 104.2±13.5% for BECKMAN COULTER Ultra Sensitive, 101.3±11.6% for CIS Bio, and 93.1±13.2% for MP kits. Sensitivity was 7.5, 6.2, 5.7, 6.2, 5.3, 4.5, and 5.5 pg/mL for SIEMENS, DIAsource, AMP, BECKMAN COULTER, BECKMAN COULTER Ultra Sensitive, CIS Bio, and MP kits, respectively. The measurement by MP kit was slightly higher than those by other kits in low-level samples, and the measurement by E170 was slightly higher than those of other kits in medium and high-level samples. In the measurement of standard sample for external quality control, SIEMENS kit produced relatively lower values whereas E170, Architect, and MP kits produced relatively higher values compared with other kits. Conclusion All tested kits for E₂ measurement have satisfactory performance for clinical use. However, correlation between kits should be considered when test kits are to be changed, because some pairs of kits do not have correlations with each other.

목차

서론
대상 및 방법
결과
고찰
결론
Acknowledgement
요약
참고문헌

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

최성희(Sung Hee Choi),노경운(Gyeong Woon Noh),김진의(Jin Eui Kim),송요성(Yoo Sung Song),팽진철(Jin Chul Paeng),강건욱(Keon Wook Kang),이동수(Dong Soo Lee). (2015).혈중 Estradiol 농도 측정을 위한 여러 방사면역측정 검사키트 및 화학면역발광 검사법의 성능평가 및 상호비교. 핵의학기술, 19 (1), 72-80

MLA

최성희(Sung Hee Choi),노경운(Gyeong Woon Noh),김진의(Jin Eui Kim),송요성(Yoo Sung Song),팽진철(Jin Chul Paeng),강건욱(Keon Wook Kang),이동수(Dong Soo Lee). "혈중 Estradiol 농도 측정을 위한 여러 방사면역측정 검사키트 및 화학면역발광 검사법의 성능평가 및 상호비교." 핵의학기술, 19.1(2015): 72-80

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제