20게이지, 23게이지, 25게이지 유리체 망막 수술 기구의 비교

이용수 0

영문명
Comparison of 20, 23 and 25 Gauge Vitreoretinal Surgical Instruments
발행기관
대한안과학회
저자명
김대우 조성원 이태곤 김철구 김종우 이재흥 한정일,Dai Woo Kim, MD, Sung Won Cho, MD, PhD, Tae Gon Lee, MD, Chul Gu Kim, MD,Jong Woo Kim, MD, PhD, Jae Heung Lee, MD, PhD, Jung Il Han, MD, PhD
간행물 정보
『대한안과학회지』Ophthalmological Society,volume54,number8, 1236~1240쪽, 전체 5쪽
주제분류
의약학 > 기타의약학
파일형태
PDF
발행일자
2013.08.15
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose: To investigate the diameter, the infusion and aspiration rate of 20-, 23- and 25-gauge vitreous cutters, and compare the theoretical and practical intraocular pressure (IOP) change according to the height of the bottle. Methods: A vitreous cutter was disassembled to measure the actual external and internal diameter. The infusion rate was measured at the height of 40, 60, 80, 100 and 120 cm and the suction rate was measured at a pressure of 100, 200, 300, 400 and 500 mm Hg for each gauge. IOP during the operation was calculated according to the bottle height and the actual IOP was measured using a mercury sphygmomanometer. Results: The external diameter of the 20-, 23- and 25-gauge was 900, 596, and 500 mm, respectively, and the internal diameter was 670, 450, and 380 mm, respectively. The infusion rate increased in direct proportion to the bottle height. The aspiration rate increased as the pressure increased. However, the increment of the infusion and aspiration rate in the 20-gauge vitreous cutter was the highest and decreased in the 23- and 25-gauge, in that order. IOP was calculated as 29, 44, 59 and 74 mm Hg when the bottle height was at 40, 60, 80 and 100 cm, respectively, and was measured 34, 50, 62 and 74 mm Hg, respectively. Conclusions: The infusion and aspiration rates per outer diameter in the 23- and 25-gauge vitreous cutters and infusions using cannula were lower than in the 20-gauge vitreous cutter. The difference of the infusion and aspiration rates per diameter between the 23- and 25-gauge vitreous cutters and infusions was not significant.

목차

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

김대우,조성원,이태곤,김철구,김종우,이재흥,한정일,Dai Woo Kim, MD, Sung Won Cho, MD, PhD, Tae Gon Lee, MD, Chul Gu Kim, MD,Jong Woo Kim, MD, PhD, Jae Heung Lee, MD, PhD, Jung Il Han, MD, PhD. (2013).20게이지, 23게이지, 25게이지 유리체 망막 수술 기구의 비교. 대한안과학회지, 54 (8), 1236-1240

MLA

김대우,조성원,이태곤,김철구,김종우,이재흥,한정일,Dai Woo Kim, MD, Sung Won Cho, MD, PhD, Tae Gon Lee, MD, Chul Gu Kim, MD,Jong Woo Kim, MD, PhD, Jae Heung Lee, MD, PhD, Jung Il Han, MD, PhD. "20게이지, 23게이지, 25게이지 유리체 망막 수술 기구의 비교." 대한안과학회지, 54.8(2013): 1236-1240

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제