학술논문
PET-CT Normalization, Well Counter Correction에 따른 팬텀을 이용한 영상 평가
이용수 2
- 영문명
- Evaluation of Image for Phantom according to Normalization, Well Counter Correction in PET-CT
- 발행기관
- 대한핵의학기술학회
- 저자명
- 이충운(Choong-Woon Lee) 유연욱(Yeon-Wook You) 문종운(Jong-Woon Mun) 김윤철(Yun-Cheol Kim)
- 간행물 정보
- 『핵의학기술』Vol.27 No.1, 37~46쪽, 전체 10쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2023.05.30
4,000원
구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.
국문 초록
영문 초록
Purpose: PET-CT imaging require an appropriate quality assurance system to achieve high efficiency and reliability. Quality control is essential for improving the quality of care and patient safety. Currently, there are performance evaluation methods of UN2-1994 and UN2-2001 proposed by NEMA and IEC for PET-CT image evaluation. In this study, we compare phantom images with the same experiments before and after PET-CT 3D normalization and well counter correction and evaluate the usefulness of quality control.
Materials and methods: Discovery 690 (General Electric Healthcare, USA) PET-CT equiptment was used to perform 3D normalization and well counter correction as recommended by GE Healthcare. Based on the recovery coefficients for the six spheres of the NEMA IEC Body Phantom recommended by the EARL. 20kBq/㎖ of 18F was injected into the sphere of the phantom and 2kBq/㎖ of 18F was injected into the body of phantom. PET-CT scan was performed with a radioacitivity ratio of 10:1. Images were reconstructed by appliying TOF+PSF+TOF, OSEM+PSF, OSEM and Gaussian filter 4.0, 4.5, 5.0, 5.5, 6.0, 6,5 ㎜ with matrix size 128×128, slice thickness 3.75 ㎜, iteration 2, subset 16 conditions. The PET image was attenuation corrected using the CT images and analyzed using software program AW 4.7 (General Electric Healthcare, USA). The ROI was set to fit 6 spheres in the CT image, RC (Recovery Coefficient) was measured after fusion of PET and CT. Statistical analysis was performed wilcoxon signed rank test using R. Results: Overall, after the quality control items were performed, the recovery coefficient of the phantom image increased and measured. Recovery coefficient according to the image reconstruction increased in the order TOF+PSF, TOF, OSEM+PSF, before and after quality control, RCmax increased by OSEM 0.13, OSEM+PSF 0.16, TOF 0.16, TOF+PSF 0.15 and RCmean increased by OSEM 0.09, OSEM+PSF 0.09, TOF 0.106, TOF+PSF 0.10. Both groups showed a statistically significant difference in wilcoxon signed rank test results (P value<0.001).
Conclusion: PET-CT system require quality assurance to achieve high efficiency and reliability. Standardized intervals and procedures should be followed for quality control. We hope that this study will be a good opportunity to think about the importance of quality control in PET-CT
목차
서론
대상 및 방법
결과
고찰
결론
REFERENCES
해당간행물 수록 논문
- 개인선량계 분실 사례 및 분실 방지 방안 조사
- 시약 변경 시 모집단 분포 추이에 따른 결과 분석
- Femoral Metastasis in Bone Mineral Densitometry
- Femoral Cystic Lesion in Bone Mineral Densitometry
- Splenic Absorption of Radiopharmaceutical in Systemic Bone Scans Performed after Liver Transplantation
- 의료기기의 발전을 위한 탄소소재의 활용
- 채혈 시간에 따른 부신피질 자극 호르몬 검사의 참고치 설정에 관한 고찰
- 감마카메라의 중장기 외인성 균일도 분석
- 대뇌 신경 교종 18F-FDG PET/CT 검사에서 포도당 부하 기법에 대한 고찰
- PET-CT Normalization, Well Counter Correction에 따른 팬텀을 이용한 영상 평가
- DXA 골밀도 검사에서 방사선사가 인식하고 있어야 할 Pitfall
- 면역방사계수측정법의 민감도 향상을 위한 변법의 평가
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 민간의료보험 해약 의향 영향 요인: 한국 의료 패널 2019년 2기 자료를 이용하여
- 의료용 인공지능에 대한 동향 분석 및 방사선사 인식조사
- 임상에서 사용되는 온장고 온도설정 및 한계점에 대한 고찰
의약학 > 방사선과학분야 NEW
- Attention UNet 모델을 활용한 유방 초음파 영상 분할: IoU 기반 학습 및 AUC, MSE 평가
- RI 등 사용 허가 대학의 정기 검사 수검 사례: 방사선학과를 대상으로
- 삼중수소 측정 신뢰성 검증을 위한 국제숙련도시험 결과 평가
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!