국문 초록
영문 초록
Purpose: The concentration of PSA (Prostate Specific Antigen) after radical prostatectomy in prostate cancer patients is a predictor of biochemical recurrence, and the AUA (American Urological Association) is defined as biochemical recurrence when the concentration of PSA is measured at 0.2 ng/mL or more, and when the concentration is measured at 0.2 ng/mL or more at the retest. This standard is also applied our hospital. In this laboratory, the PSA reagent using IRMA (Immunoradiometric Assay) is used, and the sensitivity at a very low value was not as good as the reagent used in the department of laboratory medicine. This study aims to increase the reliability of the results by improving the precision and sensitivity of very low values. Materials and Methods: As a reagent for the study, PSA reagent using IRMA was used. As a method to improve the precision and sensitivity of very low values, a variation method on the serum volume (25 uL, 50 uL, 100 uL, 200 uL) was studied, and variation usefulness evaluation was conducted. The evaluation items were compared the results of precision, analytical sensitivity, recovery rate, dilution test, high-dose hook effect test, parallel test and very low concentration values (n = 20). Results: The validation results were displayed in the order of 25 uL, 50 uL, 100 uL, 200 uL. As the serum volume increased, it was confirmed that CV (Coefficient of Variation)(%) improved. Analytical sensitivity (ng/mL) was 0.038, 0.041, 0.017, 0.015 and recovery rate (%) was 101±3, 101±3, 99±2, 97±4. very low concentration values (ng/mL) between each volume (n=20) were 0.135±0.068, 0.076±0.050, 0.048±0.034, 0.046±0.034. and high dose hook effect appeared as the serum volume increased. Conclusion: Through the variation usefulness evaluation, it was confirmed that as the serum volume increased, the precision and sensitivity improved at very low concentration values. However, it is necessary to pay special attention to the occurrence of high-dose hook effect as the serum volume increases. In the case of tests that requires very low concentration values, it is thought that the reliability of the result will be increased if the variation method is properly used after the variation usefulness evaluation.
목차
서론
실험재료 및 방법
고찰
결론
REFERENCE
해당간행물 수록 논문
- 개인선량계 분실 사례 및 분실 방지 방안 조사
- 시약 변경 시 모집단 분포 추이에 따른 결과 분석
- Femoral Metastasis in Bone Mineral Densitometry
- Femoral Cystic Lesion in Bone Mineral Densitometry
- Splenic Absorption of Radiopharmaceutical in Systemic Bone Scans Performed after Liver Transplantation
- 의료기기의 발전을 위한 탄소소재의 활용
- 채혈 시간에 따른 부신피질 자극 호르몬 검사의 참고치 설정에 관한 고찰
- 감마카메라의 중장기 외인성 균일도 분석
- 대뇌 신경 교종 18F-FDG PET/CT 검사에서 포도당 부하 기법에 대한 고찰
- PET-CT Normalization, Well Counter Correction에 따른 팬텀을 이용한 영상 평가
- DXA 골밀도 검사에서 방사선사가 인식하고 있어야 할 Pitfall
- 면역방사계수측정법의 민감도 향상을 위한 변법의 평가
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 의료용 인공지능에 대한 동향 분석 및 방사선사 인식조사
- CT contrast 검사 시 중심 정맥관을 이용한 조영제 사용에 대한 연구
- CT조영제 부작용 고위험군에 대한 전처치의 유용성
의약학 > 방사선과학분야 NEW
- 초음파 조직검사에 사용되는 Biopsy Gun Needle의 재질에 따른 반사율 연구
- 방사선에 대한 지식 및 인식도 연구
- 입원환자 일반촬영 이용량 및 피폭선량: 2018년 입원환자데이터
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!