방사성 의약품 자동 분주장치의 유용성 및 최적화된 분주방법에 관한 고찰

이용수 0

영문명
A Study on the Usefulness of Auto Dispenser and Optimized Dispensing Method
발행기관
대한핵의학기술학회
저자명
이정은(JeoungEun Lee) 김호성(Hosung Kim) 류재광(Jaekwang Ryu) 정우영(Wooyoung Jung)
간행물 정보
『핵의학기술』Vol.17 No.2, 59~66쪽, 전체 8쪽
주제분류
의약학 > 방사선과학
파일형태
PDF
발행일자
2013.11.30
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose: Recently, there is an increase of the number of hospitals using auto dispenser to reduce occupational radiation exposure when drawing up of the ¹⁸F-FDG dose (5.18 MBq/kg) in a syringe from the dramatic high activity of ¹⁸F-FDG multidose vial. The aim of this study is to confirm that using auto dispenser actually reduces the radiation exposure for technologists. Also we analyzed the reproducibility of auto dispenser to find optimized dispensing method for the device. Materials and Methods: We conducted three experiments. Comparison of radiation exposure on chest and hands: The chest and hands exposure dose received by technologists during the injection were measured by electronic personal dosimeter (EPD) and ring TLD respectively. Reproducibility of dispensed volume: We draw up the normal saline into 5 and 2 mL syringe using auto dispenser by changing the volume from 1 to 15 mm for 5 times in the same setting of the volume. The weight of 5 normal saline dispensed from the device at same volume was measured using micro balance and calculated standard deviation and coefficient of variation. Reproducibility of dispensed radioactivity: We dispensed 362.6 MBq±10%of ¹⁸F-FDG in 5 and 2 mL syringes from the multidose vial of different specific activity. In the same setting of volume, we repeated dispensing for 4 times and compared standard deviation and coefficient of variation of radioactivity between 5 syringes. Results: There was significant difference in the average of chest exposure dose according to the dispensing methods (P<0.05). Also, when dispensing ¹⁸F-FDG in manual method, exposure dose was 11.5 times higher in right hand and 4.8 times higher in left hand than in auto method. In the result of reproducibility of dispensed volume, standard deviation and coefficient of variation shows decline as the dispensing volume increases. As a result of reproducibility of dispensed radioactivity, standard deviation and coefficient of variation increases as the specific activity increases. Conclusion: We approved that the occupational radiation exposure dose of technologists were reduced when dispensing ¹⁸F-FDG using auto dose dispenser. Secondly, using small syringes helps to increase reproducibility of auto dose dispense. And also, if you lower the specific activity of ¹⁸F-FDG in multidose vial below 915-1,020 MBq/mL, you can use auto dispenser more effectively keeping the coefficient of variation lower than 10%.

목차

서론
실험재료 및 방법
결과
고찰
결론
요약
REFERENCES

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

이정은(JeoungEun Lee),김호성(Hosung Kim),류재광(Jaekwang Ryu),정우영(Wooyoung Jung). (2013).방사성 의약품 자동 분주장치의 유용성 및 최적화된 분주방법에 관한 고찰. 핵의학기술, 17 (2), 59-66

MLA

이정은(JeoungEun Lee),김호성(Hosung Kim),류재광(Jaekwang Ryu),정우영(Wooyoung Jung). "방사성 의약품 자동 분주장치의 유용성 및 최적화된 분주방법에 관한 고찰." 핵의학기술, 17.2(2013): 59-66

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제