본문 바로가기

추천 검색어

실시간 인기 검색어

학술논문

동적신장팬텀시스템 개발에 따른 장비별 사구체여과율의 비교

이용수 0

영문명
Comparison for Glomerular Filtration Rate in Gamma Camera Systems Using Dynamic Renal Phantom System
발행기관
대한핵의학기술학회
저자명
강천구(Chun Goo Kang) 박훈희(Hoon-Hee Park) 오신현(Shin Hyun Oh) 이한울(Han Wool Lee) 김정열(Jung Yul Kim) 오주영(Joo Yung Oh) 이주영(Ju Young Lee) 김재삼(Jae Sam Kim) 이창호(Chang Ho Lee)
간행물 정보
『핵의학기술』Vol.17 No.2, 3~9쪽, 전체 7쪽
주제분류
의약학 > 방사선과학
파일형태
PDF
발행일자
2013.11.30
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose: Currently commercially available phantom can reproduce and evaluate only a static situation, the study is incomplete research on phantom and system which is can confirmed functional situation in the kidney by time through dynamic phantom and blood flow velocity, various difference according to the amount of radioactive. Therefore, through this study, it has produced the dynamic kidney phantom to reproduce images through the dynamic flow of the kidney, it desire to evaluate the usefulness of nuclear medicine imaging. Materials and Methods: The production of the kidney phantom was fabricated based on the normal adult kidney, in order to reproduce the dynamic situation based on the fabricated kidney phantom, in this study it was applied the volume pump that can adjust the speed of blood flow, so it can be integrated continuously radioactive isotopes in the kidney by using 99mTc-pertechnate.Used the radioactive isotope was supplied through the two pump. It was confirmed the changes according to the infusion rate, radioactive isotopes and the different injection speeds on the left and right, analysis of the acquired images was done by drawn ten times ROI in order to check the reproducibility of each on the front and rear of the kidney and bladder. Results: Under the same conditions infusion rate 40 mL/min fixed to adjust the pressure of the pump when the radiopharmaceuticals between 2-3 minutes in the most integrated in the kidney phantom was excreted inthe bladder. Glomerular filtration rate (GFR), respectively, by each device SYMBIA 1,091 mL/min, FORTE 1,232 mL/min, ARGUS 1,264 mL/min, INFINIA 1,302 mL/min in that there isno statistically significant difference was found, Tmax values and T1/2 values stars from all equipment with no statistically significant difference was found. CV values of the coefficient of variation less than 5% was found to be repeatable, and to 2.67% of the lowest SYMBIA appeared, INFINIA was the highest in the 4.86%. Conclusion: Through this study, the results showed that the dynamic kidney phantom system is able to similarly reproduce renogram in the actual clinical. Especially, the depicted over time for the flow to be excreted through the kidney into the bladder was adequately reproduce, it is expected to be utilized as basic data to check the quality of the dynamic images. In addition, it is considered to help in the field of functional imaging and quality control.

목차

서론
대상 및 방법
결과
결론 및 고찰
요약
REFERENCES

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

강천구(Chun Goo Kang),박훈희(Hoon-Hee Park),오신현(Shin Hyun Oh),이한울(Han Wool Lee),김정열(Jung Yul Kim),오주영(Joo Yung Oh),이주영(Ju Young Lee),김재삼(Jae Sam Kim),이창호(Chang Ho Lee). (2013).동적신장팬텀시스템 개발에 따른 장비별 사구체여과율의 비교. 핵의학기술, 17 (2), 3-9

MLA

강천구(Chun Goo Kang),박훈희(Hoon-Hee Park),오신현(Shin Hyun Oh),이한울(Han Wool Lee),김정열(Jung Yul Kim),오주영(Joo Yung Oh),이주영(Ju Young Lee),김재삼(Jae Sam Kim),이창호(Chang Ho Lee). "동적신장팬텀시스템 개발에 따른 장비별 사구체여과율의 비교." 핵의학기술, 17.2(2013): 3-9

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제