감마카메라의 Sensitivity 보정 Factor에 관한 연구

이용수 0

영문명
The Correction Factor of Sensitivity in Gamma Camera: Based on Whole Body Bone Scan Image
발행기관
대한핵의학기술학회
저자명
정은미(Eun Mi Jung) 정우영(Woo Young Jung) 류재광(Jae Kwang Ryu) 김동석(Dong Seok Kim)
간행물 정보
『핵의학기술』Vol.12 No.2, 208~213쪽, 전체 6쪽
주제분류
의약학 > 방사선과학
파일형태
PDF
발행일자
2008.10.30
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose: Generally a whole body bone scan has been known as one of the most frequently executed exams in the nuclear medicine fields. Asan medical center, usually use various gamma camera systems - manufactured by PHILIPS (PRECEDENCE, BRIGHTVIEW), SIEMENS (ECAM, ECAM signature, ECAM plus, SYMBIA T2), GE (INFINIA) - to execute whole body scan. But, as we know, each camera’s sensitivity is not same so it is hard to consistent diagnosis of patients. So our purpose is when we execute whole body bone scans, we exclude uncontrollable factors and try to correct controllable factors such as inherent sensitivity of gamma camera. In this study, we’re going to measure each gamma camera’s sensitivity and study about reasonable correction factors of whole body bone scan to follow up patient’s condition using different gamma cameras. Materials and Methods: We used the ⁹⁹ᵐTc flood phantom, it recommend by IAEA recommendation based on general counts rate of a whole body scan and measured counts rates by the use of various gamma cameras - PRECEDENCE, BRIGHTVIEW, ECAM, ECAM signature, ECAM plus, IFINIA - in Asan medical center nuclear medicine department. For measuring sensitivity, all gamma camera equipped LEHR collimator (Low Energy High Resolution multi parallel Collimator) and the ⁹⁹ᵐTc gamma spectrum was adjusted around 15% window level, the photo peak was set to 140-kev and acquirded for 60 sec and 120 sec in all gamma cameras. In order to verify whether can apply calculated correction factors to whole body bone scan or not, we actually conducted the whole body bone scan to 27 patients and we compared it analyzed that results. Results: After experimenting using ⁹⁹ᵐTc flood phantom, sensitivity of ECAM plus was highest and other sensitivity order of all gamma camera is ECAM signature, SYMBIA T2, ECAM, BRIGHTVIEW, IFINIA, PRECEDENCE. And yield sensitivity correction factor show each gamma camera’s relative sensitivity ratio by yielded based on ECAM’s sensitivity. (ECAM plus 1.07, ECAM signature 1.05, SYMBIA T2 1.03, ECAM 1.00, BRIGHTVIEW 0.90, INFINIA 0.83, PRECEDENCE 0.72) When analyzing the correction factor yielded by ⁹⁹ᵐTc experiment and another correction factor yielded by whole body bone scan, it shows statistically insignificant value (p<0.05) in whole body bone scan diagnosis. Conclusion: In diagnosing the bone metastasis of patients undergoing cancer, whole body bone scan has been conducted as follow up tests due to its good points (high sensitivity, non invasive, easily conducted). But as a follow up study, it’s hard to perform whole body bone scan continuously using same gamma camera. If we use same gamma camera to patients, we have to consider effectiveness of equipment’s change by time elapsed. So we expect that applying sensitivity correction factor to patients who tested whole body bone scan regularly will add consistence in diagnosis of patients.

목차

서론
실험재료 및 방법
결과
고찰
결론
요약
REFERENCES

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

정은미(Eun Mi Jung),정우영(Woo Young Jung),류재광(Jae Kwang Ryu),김동석(Dong Seok Kim). (2008).감마카메라의 Sensitivity 보정 Factor에 관한 연구. 핵의학기술, 12 (2), 208-213

MLA

정은미(Eun Mi Jung),정우영(Woo Young Jung),류재광(Jae Kwang Ryu),김동석(Dong Seok Kim). "감마카메라의 Sensitivity 보정 Factor에 관한 연구." 핵의학기술, 12.2(2008): 208-213

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제