흉부 CT검사시 AEC Program 적용에 의한 선량감소효과에 관한 연구
이용수 45
- 영문명
- Dose Reduction Effect with Automatic Exposure Control Program in Chest CT
- 발행기관
- 대한CT영상기술학회
- 저자명
- 이대규(Dae kyoo Lee) 남윤철(Yoon Chul Nam) 김문찬(Moon Chan Kim)
- 간행물 정보
- 『대한CT영상기술학회지』대한전산화단층기술학회지 제8권 제1호, 15~23쪽, 전체 9쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2006.04.30
4,000원
구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.
국문 초록
영문 초록
Purpose
The purpose of this study is to check how much a radiation dose of the thoraxis reduced when using AEC technique in CT examination, and to apply the program properly to clinical cases.
Materials and Methods
We measured organ doses of thyroid, upper chest, mid chest, lower chest, breast, liver when conducting a test by using Brilliance CT made by Philips, instead of AEC program, and compared the radiation dose reduction effect of each organ by measuring the effect shown when using D-DOM (a method of controlling a tube current according to the thickness of a subject for photography toward X-Y axis of a revolving angle of a tube) among AEC programs, and Z-DOM(a method of controlling a tube current according to the thickness of Z-axis position).
Results
The comparison between dose with AEC program and dose without AEC program in chest routine CT were as follows,
1. The thyroid dose using D-DOM and Z-DOM wasreduced 23.9% and 49.5% respectively rather than dose of non-AEC program.
2. The upper chest dose using D-DOM and Z-DOM was reduced 18.3% and 17.8% respectively rather than dose of non-AEC program. and The mid. chestdose using D-DOM and Z-DOM was reduced 15.2% and 12.7% respectively rather than dose of non-AEC program. and The lower chest dose using D-DOM and Z-DOM was reduced 5.8% and 22.4% respectively rather than dose of non-AEC program.
3. The breast dose using D-DOM and Z-DOM was reduced 10.7% and 7.5% respectively rather than dose of non-AEC program.
4. The liver dose using D-DOM and Z-DOM was reduced 20.3% and 26.4% respectively rather than dose of non-AEC program.
5. It was proven that as for the radiation dose reduction effect by D-DOM, the control of a tube current was mostly performed in the asymmetrical part such as shoulder when comparing upper chest (18.3%), mid. Chest(15.2%), and lower chest(5.8%).
6. Thyroid, Lower chest showed a greater radiation dose reduction effect when using Z-DOM rather than D-DOM.
Conclusion
The radiation dose reduction effect in each organ was proven higher by far when using AEC program. In particular, D-DOM demonstrated an excellent radiation dose reduction effect in the part that showed a great difference of thickness between AP and Lateral(ex. shoulder). Z-DOM displayed an outstanding effect in the neck part.
목차
Abstract
Ⅰ. 목적
Ⅱ. 대상 및 방법
Ⅲ. 결과
Ⅳ. 고찰
Ⅴ. 결론
참고문헌
해당간행물 수록 논문
- Cardiac Functional Analysis를 위한 ECG-gated MDCT의 유용성과 MRI, SPECT, Echocardiography와의 비교분석
- 고해상능 전산화단층촬영을 이용한 폐기종 정량분석의 유용성
- 석회화된 관상동맥 질환에서 16 Slice MDCT의 Kernel 값이 관상동맥 내경협착에 미치는 영향평가에 대한 연구
- 전산화단층촬영장치의 품질관리 자동화 프로그램의 개발 및 적용에 관한 연구
- TLD법과 Monte Carlo Simulation에 의한 CT Organ Dose의 비교
- CT검사시 산란선에 의한 방사선피폭
- Lung Care의 Nodule Enhance View를 이용한 폐결절 발견의 유용성
- 흉부 CT검사시 AEC Program 적용에 의한 선량감소효과에 관한 연구
- CT를 이용한 복부지방 측정
- CT검사의 방사선 피폭에 관한 연구
- 테이블 이동간격 정확도와 위치확인 영상의 정확도 시험용 팬텀의 제작 및 유용성
- 하지 심부정맥혈전증에서 양측 하지 MDCT Direct Venography의 유용성
- 64 Channel Coronary Angio CT에 있어서 조영제와 Saline의 Bi-phase Injection과 재구성방법에 관한 연구
- CT Backup Server의 구성과 운용
- 자발성 뇌출혈 환자의 발생빈도
- 관상동맥 석회화 침착 정량화 CT검사를 위한 제조사별 소프트웨어의 재현성 평가
- 선량감소효과를 위한 저관전압 복부CT의 추적검사에의 적용
- 조영제 부작용 분석
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 의료용 인공지능에 대한 동향 분석 및 방사선사 인식조사
- CT contrast 검사 시 중심 정맥관을 이용한 조영제 사용에 대한 연구
- CT조영제 부작용 고위험군에 대한 전처치의 유용성
의약학 > 방사선과학분야 NEW
- 초음파 조직검사에 사용되는 Biopsy Gun Needle의 재질에 따른 반사율 연구
- 방사선에 대한 지식 및 인식도 연구
- 입원환자 일반촬영 이용량 및 피폭선량: 2018년 입원환자데이터
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!