본문 바로가기

추천 검색어

실시간 인기 검색어

학술논문

선량감소효과를 위한 저관전압 복부CT의 추적검사에의 적용

이용수 40

영문명
Low-voltage Abdominal CT for Decreasing Radiation Dose : Application to Follow-up Examination
발행기관
대한CT영상기술학회
저자명
안수현(Su Hyeon Ahn) 조인완(In Wan Cho) 강성진(Seong Jin Kang) 강래욱(Rae Wook Kang) 유병헌(Byeong Heon Yoo) 김승식(Seung Sik Kim)
간행물 정보
『대한CT영상기술학회지』대한전산화단층기술학회지 제8권 제1호, 153~160쪽, 전체 8쪽
주제분류
의약학 > 방사선과학
파일형태
PDF
발행일자
2006.04.30
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose To study the analyse radiation dose, CT number, contrast scale, image quality of the patient who needs periodic follow-up examination by applying abdominal CT using low tube voltage. Materials & Methods To evaluate the image quality of abdomen, radiation dose, CT number and contrast scale follows in change of tube voltage, subjects are divided into two groups. One group is the patients able to be applied in the clinical analysis and the other group vvere used phantom. Somatom Sensation 16(Siemens, Erlangen, Germany) was used. Twenty patient were selected at random in the group which was applied periodic follow-up examination by CT. In this clinical study, tube current was fixed in 150mA, tube voltage was adjusted according to appropriate value of each examinations and then we measured CTDI values. The measurement of CT number of each organs was performed by setting up a lcm diameter ROI at abdominal organs in the same height of first lumbar vertebra by using images of arterial phase. Quality of the abdominal images were graded into the 3 degrees by two radiologists in consensus. Abdominal shaped acrylic phantom was used in phantom study. Ion chamber was inserted into the holes located on the center and periphery of the phantom, and around there, then scaned, and radiation dose were automatically displayed by the reader which were 80kVp, 100kVp, 120kVp, 140kVp of tube voltage were applied to the phantom(diluted contrast medium with water 1:10 ratio), then the phantom was scanned and a 1cm diameter ROI in the center of the image was measured CT number. Results CTDI value was m decreased in 100kVp by 36%(11.70mGy vs 7.50mGy) in comparison with 120kVp. According to radiologist’ evaluation, there were 17 equivalent, 3 acceptable, 0 unacceptable in the group of 20 subjects. Radiation dose in the phantom study showed decreasing as tube voltage was lowered from 80kVp to 140kVp. Peripheral dose decreased by 38%, and central dose decreased by 41%. CT number showed 1365.9±4.4HU in 80kVp, 1046.1±3.7 in 100kVp, 862.8±3.2HU in 120kVp, 737.5±3.0 in 140kVp. Conclusion By applying the low tube voltage using abdominal CT to the patients who need periodical follow-up examination, the quality of the image is relatively high contrast scale without a lowering of image quality and we can decrease the effect of radiation dose.

목차

Abstract
Ⅰ. 목적
Ⅱ. 대상 및 방법
Ⅲ. 결과
Ⅳ. 고찰
Ⅴ. 결론
참고문헌

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

안수현(Su Hyeon Ahn),조인완(In Wan Cho),강성진(Seong Jin Kang),강래욱(Rae Wook Kang),유병헌(Byeong Heon Yoo),김승식(Seung Sik Kim). (2006).선량감소효과를 위한 저관전압 복부CT의 추적검사에의 적용. 대한CT영상기술학회지, 8 (1), 153-160

MLA

안수현(Su Hyeon Ahn),조인완(In Wan Cho),강성진(Seong Jin Kang),강래욱(Rae Wook Kang),유병헌(Byeong Heon Yoo),김승식(Seung Sik Kim). "선량감소효과를 위한 저관전압 복부CT의 추적검사에의 적용." 대한CT영상기술학회지, 8.1(2006): 153-160

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제