DR Mammography에서 평균유선선량 감소를 위한 방안
이용수 14
- 영문명
- The Study for Average Glandular Dose Reduction in Digital Mammography
- 발행기관
- 대한영상의학기술학회
- 저자명
- 조은혜(Eun Hye Cho) 이은주(Eun Joo Lee) 주미화(Mi Hwa Joo) 이관섭(Kwan Sup Lee) 하동윤(Dong Yoon Ha)
- 간행물 정보
- 『대한영상의학기술학회 논문지』2010년 춘계학술대회 논문지, 19~28쪽, 전체 10쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2010.03.30
4,000원
구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.
국문 초록
영문 초록
Purpose : In digital mammography, Exposure condition was automatically chosen using by measurement breast thickness and the density of mammary gland. It may cause a lack of recognition about an appropriate exposure condition and total exposure dose which would be high through the more increase dose and the image quality was better. For search a minimum dose at diagnosis and find a minimum scan parameter, in our study using an authorization standard breast phantom from ACR(American collage of radiology).
Materials and methods : AOP mode of Digital mammography(Senographe DS) used with Standard, Contrast, Dose mode take every 3 times and archived 9 images for reproducibility in same parameters. In Manual mode, kVp was fixed 29 and mAs was changed from 40 to 4 in 11steps, it take every 3 times and archived 33 images for reproducibility in same parameters. Phantom image was evaluated and analze by SPSS program with mAs, phantom scoring, evaluation of false region, image quality, SNR and average glandular dose.
Results : Average glandular dose is revealed the highest dose in contrast mode, the lowest dose in dose mode out of AOP mode.
The scoring of false region was passed on ACR quality control standard, it was above 0.58mGy at average glandular dose(59% of standard mode) and 25mAs(61% of standard mode). It has a right statistical correlation. Do evaluations of false region with the naked eye, the order of correlation with average glandular dose was such as following speck, fiber, mass. It has a right statistical correlation. Also evaluation of image quality was above score 3(good) which more than 0.92mGy(93% of standard mode) and 40mAs(98% of standard mode). It has a right statistical correlation. SNR was satisfied with standard score, more than 0.58mGy average glandular dose(63% of standard mode), and 25mAs(61% of standard mode). It has a right statistical correlation, too.
Conclusion : The image quality of possibly diagnosis could maintain by low average glandular dose and mAs by Manual mode which less than AOP mode parameters of the existing digital mammography. Also it could reduce average glandular dose. Also it could change minimum parameters accompany with a shape of lesion. This method is innovative plan of reduction of patient dose about unnecessary radiation exposure by using a AOP mode but Manual mode with previous reference parameters when follow-up study and post operation patient exam.
목차
Abstact
Ⅰ. 서론
Ⅱ. 대상 및 방법
Ⅲ. 결과
Ⅳ. 고찰
Ⅴ. 결론
참고문헌
해당간행물 수록 논문
- 담관 결석 발생과 재발에 있어 십이지장 팽대부 주위 게실과의 상관관계
- Source Image Distance 변화에 따른 환자 및 검사자의 피폭선량 비교
- 유방촬영술에서 방사선 반 투과성(Radiolucent) 패드(Mammo Pad)의 유용성 관한 고찰
- 수술 중 X-ray 검사 시 Cassette Draping에 대한 고찰
- 방사선투시장비를 사용한 검사에 참여하는 소화기내과 소속 간호사들의 의료용 방사선에 대한 의식도 조사
- 폐경기 여성에서 DXA를 이용한 Lateral Vertebral Assessment의 유용성
- 조기크론병 환자의 진단에 있어서 임상양상과 성격유형의 비교분석에 따른 Peroral Pneumocolon Technique의 유용성
- 선원에서 X선 방어용 칸막이간 거리가 상이한 유방촬영실의 공간 산란선량 비교
- Fracture of Hamate Hook의 유용한 방사선 단순촬영법에 대한 고찰
- 투시조영검사 시 환자 피폭선량 저감화 대책에 관한 연구
- 소아 투시조영검사 시 Pulse rate별 화질분석을 통한 저선량 모드 선택에 관한 연구
- 개선된 Fulcrum을 이용한 Bending radiograph의 유용성에 관한 고찰
- 고관절 표면 치환술 환자의 LATERAL 검사법에 대한 고찰
- Shoulder Rockwood 검사법에서 Position 변화에 따른 Image Quality 향상을 위한 연구
- DR Mammography에서 평균유선선량 감소를 위한 방안
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 의료용 인공지능에 대한 동향 분석 및 방사선사 인식조사
- CT contrast 검사 시 중심 정맥관을 이용한 조영제 사용에 대한 연구
- CT조영제 부작용 고위험군에 대한 전처치의 유용성
의약학 > 방사선과학분야 NEW
- 초음파 조직검사에 사용되는 Biopsy Gun Needle의 재질에 따른 반사율 연구
- 방사선에 대한 지식 및 인식도 연구
- 입원환자 일반촬영 이용량 및 피폭선량: 2018년 입원환자데이터
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!