A PC-Based Monte Carlo Program과 유리선량계를 이용한 장기선량 및 유효선량 평가

이용수 26

영문명
Evaluation of Organ and Effective Dose using A PC-Based Monte Carlo Program and Glass Dosimeter
발행기관
대한영상의학기술학회
저자명
최원근(Weon Keun Choi) 김재인(Jae Yin Kim) 홍준기(Joon Ki Hong) 김성진(Seong Jin Kim) 이관섭(Kwan Sub Lee) 하동윤(Dong Yoon Ha)
간행물 정보
『대한영상의학기술학회 논문지』2009년 춘계학술대회 논문지, 131~138쪽, 전체 8쪽
주제분류
의약학 > 방사선과학
파일형태
PDF
발행일자
2009.03.30
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose : The purpose of this study is to examine the utility of PCXMC by comparing organ dose based on statistical data obtained from Monte Carlo of program and organ dose obtained from the method using phantom and glass dosimeter. Materials and Methods : 1. Direct method (the method of measurement using Glass Dosimeter) : GD3.52M (GD3.52M will be referred to as element ) is inserted in the center of each organ. To calculate HVL and entrance exposure dose(mR), Unfors Xi meter is attached to around center of phantom entrance surface. To calculate ESD, five elements were attached around fields. Chest PA, Chest Lat, L-spine AP, L-spine Lat, Pelvis AP, and Skull PA are irradiated as Protocol of this hospital. 2. indirect method (the method using PCXMC program) : PCXMC program minimizes Error (%) by entering the same parameter as direct method on the hypothetical mathematical phantom The same age, stature, weight, and Focus-Skin Distance as Examination data of RANDO phantom are input and the FOV is drawn in the very same position. 16 organ doses are compared and analyzed which are the same as glass dosimeter among 29 organ doses obtained from inputting doses in PCXMC. Results : 1. Organ doses : In all six tests, when comparing between PCXMC within FOV and others with organ doses in glass dosimeter, they are significant with each other (p<0.05). With relation to Out of FOV organ dose of Chest PA, Chest Lat, Pelvis AP, and Skull PA, the difference in terms of statistic is significant (p<0.05). 2. Effective dose : In Chest PA, and Chest Lat, effective dose using PCXMC is 0.0453, 0.1542 and effective dose using glass dosimeter is 0.0527, 0.1699 and shows the difference of 14% and 9% respectively. In Skull PA, organ dose using PCXMC program is 0.0209 and organic dose using glass dosimeter is 0.0209 and therefore they show the same result. In L-spine AP, and Lat, effective dose using PCXMC is 0.5635, 0.4794 and effective dose using glass dosimeter is 0.7845, 0.6301 and in other words, 28%, 23% respectively. The method of measurement using glass dosimeter is higher than that of measurement using PCXMC. In Pelvis AP, effective dose using PCXMC is 0.4709 and effective dose using glass dosimeter is 0.6833 and the difference is 31%. Conclusion : 1. Organ dose : In cases of organ dose using glass dosimeter and organ dose using PCXMC program and Chest PA, Chest Lat, Pelvis AP and Skull PA in which there is the difference of average, they are organ dose out of FOV and low dose. In case predose measurement value is less than 30uGy of organ dose, there is difference in measurement value in between organ dose using PCXMC program and organic dose using glass dosimeter. 2. Effective dose : In Chest PA, Lat, and Skull PA which show low effective dose, effective dose using PCXMC program and effective dose using glass dosimeter show similar result. In L-spine, Lat, and Pelvis AP with high effective, effective dose using glass dosimeter compared with effective dose using PCXMC program is measured high consistently.

목차

Abstact
Ⅰ. 서론
Ⅱ. 대상 및 방법
Ⅲ. 결과
Ⅳ. 고찰
Ⅴ. 결론
참고문헌

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

최원근(Weon Keun Choi),김재인(Jae Yin Kim),홍준기(Joon Ki Hong),김성진(Seong Jin Kim),이관섭(Kwan Sub Lee),하동윤(Dong Yoon Ha). (2009).A PC-Based Monte Carlo Program과 유리선량계를 이용한 장기선량 및 유효선량 평가. 대한영상의학기술학회 논문지, 2009 (1), 131-138

MLA

최원근(Weon Keun Choi),김재인(Jae Yin Kim),홍준기(Joon Ki Hong),김성진(Seong Jin Kim),이관섭(Kwan Sub Lee),하동윤(Dong Yoon Ha). "A PC-Based Monte Carlo Program과 유리선량계를 이용한 장기선량 및 유효선량 평가." 대한영상의학기술학회 논문지, 2009.1(2009): 131-138

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제