선형가속기를 이용한 방사선 수술의 선량분포의 실험적 확인

이용수 6

영문명
Verification of Dose Distribution for Stereotactic Radiosurgery with a Linear Accelerator
발행기관
대한방사선종양학회
저자명
박경란(Kyung Ran Park) 김계준(Kye Jun Kim) 추성실(Sung Sil Chu) 이종영(Jong Young Lee) 조철우(Chul Woo Joh) 이창걸(Chang Geol Lee) 서창옥(Chang Ok Suh) 김귀언(Gwi Eon Kim)
간행물 정보
『대한방사선종양학회지』제11권 제2호, 421~430쪽, 전체 10쪽
주제분류
의약학 > 종양학
파일형태
PDF
발행일자
1993.12.29
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

The calculation of dose distribution in multiple arc stereotactic radiotherapy is a three-dimensional problem and, therefore, the three-dimensional dose calculation algorithm is important and the algorithm's accuracy and reliability should be confirmed experimentally. The aim of this study is to verify the dose distribution of stereotactic radiosurgery experimentally and to investigate the effect of the beam quality, the number of arcs of radiation, and the tertiary collimation on the resulting dose distribution. Film dosimetry with phantom measurements was done to get the three-dimensional orthogonal isodose distribution. All experiments were carried out with a 6 MV X-ray, except for the study of the effects of beam energy on dose distribution, which was done for X-ray energies of 6 and 15 MV. The irradiation technique was from 4 to 11 arcs at intervals of from 15 to 45 degrees between each arc with various field sizes with additional circular collimator. The dose distributions of square field with linear accelerator collimator compared with the dose distributions obtained using circular field with tertiary collimator. The parameters used for comparing the results were the shape of the isodose curve, dose fall-offs fom 90% to 50% and from 90%;to;20% isodose line for the steepest and shallowest profile, and A={90% isodose area / (50% isodose area - 90% isodose area)}(modified from Chierego). This ratio may be considered as being proportional to the sparing of normal tissue around the target volume. The effect of beam energy in 6 and 15 MV X-ray indicated that the shapes of isodose curves were the same. The value of ratio A and the steepest and shallowest dose fall-offs for 6 MV X-ray was minimally better than that for 15 MV X-ray. These data illustrated that an increase in the dimensions of the field from 10 to 28 mm in diameter did not significantly change the isodose distribution. There was no significant difference in dose gradient and the shape of isodose curve regardless of the number of arcs for field sizes of 10, 21, and 32 mm in diameter. The shape of isodose curves was more circular in circular field and square in square field. And the dose gradient for the circular field was slightly better than that for the square field.

목차

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

박경란(Kyung Ran Park),김계준(Kye Jun Kim),추성실(Sung Sil Chu),이종영(Jong Young Lee),조철우(Chul Woo Joh),이창걸(Chang Geol Lee),서창옥(Chang Ok Suh),김귀언(Gwi Eon Kim). (1993).선형가속기를 이용한 방사선 수술의 선량분포의 실험적 확인. 대한방사선종양학회지, 11 (2), 421-430

MLA

박경란(Kyung Ran Park),김계준(Kye Jun Kim),추성실(Sung Sil Chu),이종영(Jong Young Lee),조철우(Chul Woo Joh),이창걸(Chang Geol Lee),서창옥(Chang Ok Suh),김귀언(Gwi Eon Kim). "선형가속기를 이용한 방사선 수술의 선량분포의 실험적 확인." 대한방사선종양학회지, 11.2(1993): 421-430

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제