학술논문
Westgard Multi-Rules의 효율적 적용과 조치사항의 개선
이용수 0
- 영문명
- Efficient Application of Westgard Multi-Rules and Quality Control Implementation Improvement
- 발행기관
- 대한핵의학기술학회
- 저자명
- 정흥수(Heung Soo Jung) 오윤정(Youn Jung Oh) 배진수(Jin Soo Bae) 백진영(Jin Young Baek) 황보라(Bo ra Hwang) 신용환(Yong Hwan Shin)
- 간행물 정보
- 『핵의학기술』Vol.21 No.1, 60~64쪽, 전체 5쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2017.06.30
4,000원
구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.
국문 초록
영문 초록
Purpose Westgard multi-rules application based on test quality improvement and commercialized international standard has been widely used in quality control. However, it is difficult to applicate the Westgard multi-rules in nuclear medicine in vitro tests due to the larger sample sizes and the simultaneous measurement of quality control material and patient sample. This study investigated the usefulness of Westgard multi-rules application in nuclear medicine in vitro tests. Materials and Methods A total of 282 systematic error multi-rules (22s, 101s) recorded in the samsung medical center computer system from January 2013 to June 2016 along with 117 cases of corrective measure record was analyzed. The Quality control implementation is recorded in Hospital information system were divided into 4 high-level areas including quality control material error, experimental procedural error, Kit lot number management error, and others. To prevent quality control material error, the existing method that each staff used their own method was changed. The staff who in charge of managing the quality control material was designated and daily consumption amount of every test was strictly controlled by one person. To prevent other errors, every test step was standardized so that the entire test procedures are identically implemented. Results The total quality control implementation was 117 cases; As a result, 62 quality control material errors were 62 cases, experimental process errors were 24 cases, Kit lot number control errors were 18 cases, and other errors were 13 cases. The quality control material error was corrected and could be used fresh materials within 2 days after thawing. The cases of systemic error were decreased to causes as quality control material error. The quality control materials were reduced above 10 vials to a monthly average. In addition, these errors of experimental processing and Kit lot number were improved by test standardization. Consequently, the cases of 101s and 22s in systematic error rules decreased at least 2 cases to a monthly average. Conclusion To confirm of systematic error through multi-rules application quickly, it is necessary to base on management of the QC material, target values and standard deviation. Moreover, in the event of a systematic error, it was found important to record measures based on test cause analysis. The experiment results are expected to contribute to internal quality control improvement and prompt and accurate result reporting through error recording and causal analysis based on Westgard multi-rules analysis.
목차
서론
실험대상 및 방법
결과
고찰 및 결론
요약
REFERENCES
해당간행물 수록 논문
- 설문조사를 바탕으로 부서 간 만족도 향상을 위한 개선 안
- Thyroglobulin 검사의 반응조건에 따른 결과 비교 분석
- 방사성 동위원소를 이용한 핵의학과 검사에서 병동 간호사의 방사선 피폭선량 평가
- 대한핵의학기술학회 투고 논문 분류
- 전신 뼈 스캔의 오류 감소를 위한 프로세스 구축과 적용 효과
- Westgard Multi-Rules의 효율적 적용과 조치사항의 개선
- Brain SPECT 검사 시 Dynamic Continuous Mode의 유용성 평가
- 물리적 보정된 핵의학 영상 평가
- PET-CT 검사 시 호흡 동조 시스템들의 유용성 평가
- Thyroglobulin 검사의 반응조건에 따른 결과 비교 분석
- ⁵⁷Co 밀봉선원의 노후도에 따른 균일도의 변동과 중첩사용에 관한 유용성 평가
- PET/CT 검사에서 Flow mode를 적용한 Respiratory Gating Method 촬영과 추가 Gating 촬영의 비교 및 유용성 평가
- 갑상선 섭취율 측정에서 선원의 길이와 부피 증가에 따른 영향
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 의료용 인공지능에 대한 동향 분석 및 방사선사 인식조사
- CT contrast 검사 시 중심 정맥관을 이용한 조영제 사용에 대한 연구
- CT조영제 부작용 고위험군에 대한 전처치의 유용성
의약학 > 방사선과학분야 NEW
- 초음파 조직검사에 사용되는 Biopsy Gun Needle의 재질에 따른 반사율 연구
- 방사선에 대한 지식 및 인식도 연구
- 입원환자 일반촬영 이용량 및 피폭선량: 2018년 입원환자데이터
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)