심근 관류 SPECT에서 핵종에 따른 Uniformity correction map 설정을 통한 영상의 질 비교
이용수 0
- 영문명
- The difference of image quality using other radioactive isotope in uniformity correction map of myocardial perfusion SPECT
- 발행기관
- 대한핵의학기술학회
- 저자명
- 송재혁(Jae hyuk Song) 김경식(Kyeong Sik Kim) 이동훈(Dong Hoon Lee) 김성환(Sung Hwan Kim) 박장원(Jang Won Park)
- 간행물 정보
- 『핵의학기술』Vol.19 No.2, 87~92쪽, 전체 6쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2015.10.30
4,000원
구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.
국문 초록
영문 초록
Purpose When the patients takes myocardial perfusion SPECT using ²⁰¹Tl, the operator gives the patients an injection of ²⁰¹Tl. But the uniformity correction map in SPECT uses ⁹⁹ᵐTc uniformity correction map. Thus, we want to compare the image quality when it uses ⁹⁹ᵐTc uniformity correction map and when it uses ²⁰¹Tl uniformity correction map. Materials and Methods Phantom study is performed. We take the data by Asan medical center daily QC condition with flood phantom including ²⁰¹Tl 21.3 kBq/mL. After postprocessing with this data, we analyze CFOV integral uniformity(I.U) and differential uniformity(D.U). And we take the data with Jaszczak ECT Phantom by American college of radiology accreditation program instruction including ²⁰¹Tl 33.4 kBq/mL. After post processing with this data, we analyze spatial Resolution, Integral Uniformity(I.U), coefficient of variation(C.V) and Contrast with Interactive data language program. Results In the flood phantom test, when it uses ⁹⁹ᵐTc uniformity correction map, Flood I.U is 3.6% and D.U is 3.0%. When it uses ²⁰¹Tl uniformity correction map, Flood I.U is 3.8% and D.U is 2.1%. The flood I.U is worsen about 5%, but the D.U is improved about 30% inversely. In the Jaszczak ECT phantom test, when it uses ⁹⁹ᵐTc uniformity correction map, SPECT I.U, C.V and contrast is 13.99%, 4.89% and 0.69. When it uses ²⁰¹Tl uniformity correction map, SPECT I.U, C.V and contrast is 11.37%, 4.79% and 0.78. All of data are improved about 18%, 2%, 13% The spatial resolution was no significant changes. Conclusion In the flood phantom test, Flood I.U is worsen but Flood D.U is improved. Therefore, it’s uncertain that an image quality is improved with flood phantom test. On the other hand, SPECT I.U, C.V, Contrast are improved about 18%, 2%, 13% in the Jaszczak ECT phantom test. This study has limitations that we can’t take all variables into account and study with two phantoms. We need think about things that it has a good effect when doctors decipher the nuclear medicine image and it’s possible to improve the image quality using the uniformity correction map of other radionuclides other than ⁹⁹ᵐTc, ²⁰¹Tl when we make other nuclear medicine examinations.
목차
서론
실험방법 및 재료
결과
결론 및 고찰
요약
참고문헌
해당간행물 수록 논문
- Ultrasensitive serum thyroglobulin의 유용성 평가
- Breast lymphoscintigraphy 검사 시 체표윤곽을 나타내는 방법의 비교
- PET/CT 검사 시 CT 피폭선량 감소 방법들의 최적화 평가
- Brain PET에서 Truncated Region에 의한 영상의 질 평가
- 장비정도관리에 Calibrator[I-125] Set 적용
- 심근 관류 SPECT에서 핵종에 따른 Uniformity correction map 설정을 통한 영상의 질 비교
- SPECT/CT 영상에서 에너지창 기반 산란보정과 CT 기반 산란보정 방법의 정량적 정확성 비교
- 25 Vitamin D3 측정에 있어서 화학발광미세입자 측정법과 화학발광면역 측정법 간의 비교 및 고찰
- PET-CT 검사 시 호흡동조 시스템을 이용한 인공물 감소에 대한 비교 평가
- 동적 타액선 신티그라피에서 기능적 매개변수 지표를 활용한 정량 분석의 유용성
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 의료용 인공지능에 대한 동향 분석 및 방사선사 인식조사
- CT contrast 검사 시 중심 정맥관을 이용한 조영제 사용에 대한 연구
- CT조영제 부작용 고위험군에 대한 전처치의 유용성
의약학 > 방사선과학분야 NEW
- 초음파 조직검사에 사용되는 Biopsy Gun Needle의 재질에 따른 반사율 연구
- 방사선에 대한 지식 및 인식도 연구
- 입원환자 일반촬영 이용량 및 피폭선량: 2018년 입원환자데이터
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!