PET/CT 검사 시 CT 피폭선량 감소 방법들의 최적화 평가

이용수 0

영문명
Evaluation for Optimization of CT Dose Reduction Methods in PET/CT
발행기관
대한핵의학기술학회
저자명
도용호(Yong Ho Do) 이홍재(Hong Jae Lee) 김진의(Jin Eui Kim)
간행물 정보
『핵의학기술』Vol.19 No.2, 55~62쪽, 전체 8쪽
주제분류
의약학 > 방사선과학
파일형태
PDF
발행일자
2015.10.30
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose Various methods for reducing radiation exposure have been continuously being developed. The aim of this study is to evaluate effectiveness of dose reduction, image quality and PET SUV changes by applying combination of automatic exposure dose(AEC), automated dose-optimized selection of X-ray tube voltage(CAREkV) and sinogram affirmed iterative reconstruction(SAFIRE) which can be controled by user. Materials and Methods Torso, AAPM CT performance and IEC body phantom images were acquired using biograph mCT64, (Siemens, Germany) PET/CT scanner. Standard CT condition was 120 kV, 40 mAs. Radiation exposure and noise were evaluated by applying AEC, CAREkV(120 kV, 40 mAs) and SAFIRE(120 kV, 25 mAs) with torso phantom compare to standard CT condition. And torso, AAPM and IEC phantom images were acquired with combination of 3 methods in condition of 120 kV, 25 mAs to evaluate radiation exposure, noise, spatial resolution and SUV changes. Results When applying AEC, CTDIvol and DLP were decreased by 50.52% and 50.62% compare to images which is not applying AEC. mAs was increased by 61.5% to compensate image quality according to decreasing 20 kV when applying CAREkV. However, CTDIvol and DLP were decreased by 6.2% and 5.5%. When reference mAs was the lower and strength was the higher, reduction of radiation exposure rate was the bigger. Mean SD and DLP were decreased by 2.2% and 38% when applying SAFIRE even though mAs was decreased by 37.5%(from 40 mAs to 25 mAs). Combination of 3 methods test, SD decreased by 5.17% and there was no significant differences in spatial resolution. And mean SD and DLP were decreased by 6.7% and 36.9% compare to 120 kV, 40 mAs with AEC. For SUV test, there was no statistical differences(P>0.05). Conclusion Combination of 3 methods shows dose reduction effect without degrading image quality and SUV changes. To reduce radiation exposure in PET/CT study, continuous effort is needed by optimizing various dose reduction methods.

목차

서론
실험재료 및 방법
결과
고찰
결론
요약
참고문헌

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

도용호(Yong Ho Do),이홍재(Hong Jae Lee),김진의(Jin Eui Kim). (2015).PET/CT 검사 시 CT 피폭선량 감소 방법들의 최적화 평가. 핵의학기술, 19 (2), 55-62

MLA

도용호(Yong Ho Do),이홍재(Hong Jae Lee),김진의(Jin Eui Kim). "PET/CT 검사 시 CT 피폭선량 감소 방법들의 최적화 평가." 핵의학기술, 19.2(2015): 55-62

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제