본문 바로가기

추천 검색어

실시간 인기 검색어

학술논문

전신 뼈 검사에서 의료용 신체 고정구의 유용성 평가

이용수 5

영문명
Evaluating the Usability of Medical Body Wrap in Whole Body Bone Scan
발행기관
대한핵의학기술학회
저자명
심동오(Dong-Oh Shim) 정우영(Woo-Young Jung) 류재광(Jae-Kwang Ryu) 박철홍(Cheol-Hong Park) 김윤재(Yoon-Jae Kim)
간행물 정보
『핵의학기술』Vol.28 No.1, 49~56쪽, 전체 8쪽
주제분류
의약학 > 방사선과학
파일형태
PDF
발행일자
2024.05.30
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose: When performing nuclear medicine examinations, body wraps or plastic supports are used to support and immobilize the patient's upper extremities to prevent patient safety accidents. However, the existing plastic supports compromised patient and staff safety, including finger entrapment and falls. Moreover, the body wrap provided by manufacturers compromised image quality such as upper extremities cutoff during whole body bone scan. Therefore, a new design of body wrap was developed to improve the issue, and this study aims to evaluate the usability of this medical body wrap. Materials and Methods: To evaluate the usability of the newly designed medical body wrap, a quality assessment of whole body bone scan images and a user satisfaction survey were conducted. Adult patients (male:female=129:152, age: 60.3±12.4 years, BMI: 24.0±4.2) aged 16 years or older who underwent a whole body bone scan during two periods: June to July 2022 (before improvement, n=139) and June to July 2023 (after improvement, n=142) were randomly selected for image quality evaluation. Five radiotechnologists visually evaluated the posterior view of the whole body bone image, including the left and right elbow (2 points), arm (2 points), whether the hand is extended (2 points), whether the hand is included (2 points), and the number of visible fingers (10 points), with a total of 18 points, which were converted to 100 points and analyzed for difference before and after improvement using an independent sample t-test. The user satisfaction questionnaire was evaluated using a 5-point Likert scale among 16 radiotechnologists from three general hospitals who experienced the new body wrap. Results: The image quality assessment was 82.0±13.8 before the improvement and 89.3±10.1 after the improvement, an average of 7.3 points higher, with a statistically significant difference (t=5.02, p<0.01). The user satisfaction survey showed an overall satisfaction rating of 4.1±0.8 for ease of use, 3.8±0.7 for scan preparation time, 3.9±0.7 for patient safety, 3.8±1.2 for scan accuracy, and 4.2±0.7 for recommendation (87.5% questionnaire response rate). Conclusion: The developed body wrap showed higher image quality and user satisfaction compared to the old method. Considering these results, it is deemed that the new body wrap may be more useful than existing methods.

목차

서론
대상 및 방법
결과
고찰
결론
REFERENCES

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

심동오(Dong-Oh Shim),정우영(Woo-Young Jung),류재광(Jae-Kwang Ryu),박철홍(Cheol-Hong Park),김윤재(Yoon-Jae Kim). (2024).전신 뼈 검사에서 의료용 신체 고정구의 유용성 평가. 핵의학기술, 28 (1), 49-56

MLA

심동오(Dong-Oh Shim),정우영(Woo-Young Jung),류재광(Jae-Kwang Ryu),박철홍(Cheol-Hong Park),김윤재(Yoon-Jae Kim). "전신 뼈 검사에서 의료용 신체 고정구의 유용성 평가." 핵의학기술, 28.1(2024): 49-56

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제