하이브리드 반복 재구성 기법이 적용된 흉부 CT영상에서 결절의 체적변화 분석
이용수 30
- 영문명
- Analysis of Pulmonary Nodule’s Volume Change on Chest CT Image with Hybrid Iterative Reconstruction Technique
- 발행기관
- 대한CT영상기술학회
- 저자명
- 박상훈(Sang-Hoon Park) (Yung-Kyoon Kim) (Wan-Yuk Kim) (Kwang-Hyun Jang) (Yoon-Chul Nam) (Mi-Sub Ahn)
- 간행물 정보
- 『대한CT영상기술학회지』대한CT영상기술학회지 제20권 제2호, 107~114쪽, 전체 8쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2018.09.30
4,000원
구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.
국문 초록
목적: CT영상 재구성 기법에서 필터 보정 역투영법(이하 FBP)과 비교하여 단계별 하이브리드 반복 재구성 기법(이하Hybrid IR)을 적용하였을 때, 흉부 CT영상에서 흉부 결절 체적 값에 변화가 있는지에 대해 확인한다.
대상 및 방법: CT장비는 GEDiscovery 750 (General Electric medical systems, Milwaukee, USA) CT와Toshiba aquilion one(Toshiba Medical System Corporation, Tochigi, Japan) 장비를 사용 하였다. Lungman phantom(Kyotokagaku, japan)을 이용하였으며흉부 결절은Lungman phantom 에서 제공하는 감약계수 값 +100 결절을 사용하였으며 직경 12mm, 10mm, 8mm의 세 종류를 Lungman phantom 안쪽에 삽입하여 두 장비에서 각각 30회씩 검사한다. 이때 조건은 120kV, 300 mA, 1:1 Pitch, thickness 0.625mm, standard algorithm로 Scan하였다. 검사한 영상을GE CT영상은FBP, ASIR30%, ASIR 60%, ASIR 90% 4 단계로 영상을 재구성하며, Toshiba CT영상은 FBP, ADIR mild, ADIR standard, ADIR strong의 4단계로 영상을 재구성 한 뒤 AW volume share 5 에서 3차원 재구성 방법인 Volume Rendering 기법을 통하여 체적을측정한 뒤 통계 낸 후 비교하였다.통계는 PASW Statistics 18. Release 18.0.0과
집단의 정규성 검정을 위하여 Kolmogorov-Smirnova test를 사용하였고, 각 집단간 평균의 차이를 평가하기 위하여 일원 배치
분산 분석 하였다. 이 때 사후 검정은 scheffe분석을 사용하였다.
결과: 본 연구에서 두 회사의 CT 영상에서의 흉부 결절 체적 측정의 평균 값은 실제 체적 값보다 과대평가 되었다. GE사 장비 에서 FBP기법 시 평균 체적은 실제 체적 보다 최대 6.79%, 최소 3.77%, Hybrid IR기법에서는 최대 5.96%, 최소 1.73% 차이를 보였으며, Toshiba 장비 에서는 FBP기법 시 최대 12.42%, 최소 4.76%, Hybrid IR은최대 12.23%, 최소 4.36% 실제 체적 보다 과대 평가 하였다. 결절의 크기나 장비에 관계 없이 FBP기법으로 재구성 하였을 때 Hybrid IR 보다 체적을 더 많이 과대 평가 하였으며, Hybrid IR의 강도가 높아질수록 체적의 확대율은 점점 줄어들었으며, FBP와는 최대 2.46%의 차이를 보였다
(p<0.05).
결론: CT검사에서 Hybrid IR 강도가 커질수록 noise가 감소하여 결절의 체적이 감소하여 FBP보다 실측치에 가까워 진다. 그러나
Hybrid IR역시 결절 체적을 과대평가 하며 진료 및 판독 시 오류를 범할 수 있으므로 이에 따른 보정이 필요하다.
영문 초록
Purpose: We estimated variation with the value of chest nodules volume in Chest CT image compare to reconstruction technique between the Hybrid Iterative Reconstruction (Hybrid IR) and the Filtered Back projection (FBP).
Material and methods: We inserted three different chest nodules, which of diameters were 12mm, 10mm, and 8mm in the Lungman phantom. And phantom images were obtained 30 times and were reconstructed in 4 steps IR each by GE and Toshiba CT(120kVp, 300mA, standard algorithm). We compared after volume was measured by AW volume share 5. Statistics were obtained by using PASW Statistical 18. Release 18.0.0.
Result: From this study, the value of average was overestimated compared to the actual value in CT image from GE and Toshiba equipment. Regardless of the size of the nodule and type of equipment, the magnification was the largest with reconstructed by FBP and was overestimated up to 12.4% of the actual volume. As the Hybrid IR intensity was increased, the magnification rate of the volume gradually was decreased. There was a difference of 2.46% at maximum compared to FBP(p<0.05)
Conclusion: The volume of chest nodule obtained by image reconstruction in CT scan was overestimated than the actual value. According to decreased of noise, volume of chest nodule was closed to the actual value as the intensity of Hybrid IR was increased than the image by FBP technique. However, the Hybrid IR technique was also caused overestimation of nodule volume. So, adapted correction was required to prevent the errors of reading and treatment in Chest CT scan.
목차
I. INTRODUCTION
II. MATERIAL AND METHODS
III. RESULT
IV. DISCUSSION
V. CONCLUSION
키워드
해당간행물 수록 논문
- CT TAVI(trans-catheter aortic valve implantation) 검사에 있어 Smart coverage 기능의 유용성 평가
- 경부 CT에서 치아 임플란트에 의한 금속성 인공음영 감소의 유용성
- 반복적 재구성 기법이 관상동맥 내 석회화 수치 측정에 미치는 영향
- Paranasal sinus CT검사시 Virtual Non Contrast 기법을 통한 화질 평가 및선량감소 효과에 대한 연구
- 3D Anteversion Angle CT에서 Bone Metal Modality Preset의 진단적 유용성
- 3차원 프린팅을 위한 CT 데이터 후처리 가이드라인: 선천성 심장병(Congenital Heart Defect, CHD)
- 최저 심박수 구간을 이용한 Coronary artery CT Angiography의 화질 개선에 관한 연구
- CT 검사시 조영제 주입 후 체온변화에 대한 연구
- Thyroid with Neck CT 검사에서 저관전압 70 kVp 프로토콜 검사의 선량감소 효과와 영상의 유용성에 관한 연구
- CINE CT 검사 시 소아 기관-기관지연화증의 유용성 평가
- 이중에너지를 이용한 복부 CT 검사의 방사선량에 관한 연구
- 하이브리드 반복 재구성 기법이 적용된 흉부 CT영상에서 결절의 체적변화 분석
- Dual energy를 이용한 cardiac CT에서 virtual calcium scoring의 유용성 평가
- Metal artifact reduction algorithm과 MonoE reconstruction(Dual Layer Detector) 중복 적용 시 유용성 연구
- AI(Artificial Intelligence)를 도입한 CT Big data 기반으로 예약 Algorithm 개발에 대한 연구
- 필터보정 역투영법. 모델기반 반복적재구성법, ASIR-V로 획득한 하지정맥 혈류영상과 방사선량 분석
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 의료용 인공지능에 대한 동향 분석 및 방사선사 인식조사
- CT contrast 검사 시 중심 정맥관을 이용한 조영제 사용에 대한 연구
- CT조영제 부작용 고위험군에 대한 전처치의 유용성
의약학 > 방사선과학분야 NEW
- 초음파 조직검사에 사용되는 Biopsy Gun Needle의 재질에 따른 반사율 연구
- 방사선에 대한 지식 및 인식도 연구
- 입원환자 일반촬영 이용량 및 피폭선량: 2018년 입원환자데이터
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!