본문 바로가기

추천 검색어

실시간 인기 검색어

학술논문

Repeatability and Agreement of Chord Mu between Scheimpflug Tomography and Swept-Source Optical Coherence Tomography

이용수 0

영문명
Repeatability and Agreement of Chord Mu between Scheimpflug Tomography and Swept-Source Optical Coherence Tomography
발행기관
대한안과학회
저자명
Na Hyun Kim Hyun Jin Kim Soo Chang Cho Kyung Eun Han
간행물 정보
『The Korean Journal of Ophthalmology』Vol.37 No.6, 510~517쪽, 전체 8쪽
주제분류
의약학 > 의학일반
파일형태
PDF
발행일자
2023.12.31
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Purpose: To evaluate repeatability and agreement of chord mu between Scheimpflug tomography (Pentacam HR) and sweptsource optical coherence tomography-based optical biometer (IOLMaster 700). Methods: In this retrospective study, 63 eyes from 33 patients were included. Chord mu, X and Y Cartesian distances between the corneal vertex and the pupil center (Px and Py), and the pupil diameter were compared using two instruments. Repeatability was evaluated using intraclass correlation coefficient (ICC), coefficient of variation (CoV), and within-subject standard deviation (Sw). Interdevice agreement was evaluated using paired t-tests and Bland-Altman plots. Results: Although Sw values for all parameters were similar between the two devices, CoV values of chord mu and pupil diameter were lower, and ICC values of those parameters were higher, in the IOLMaster 700 than in the Pentacam HR. Chord mu and pupil diameter values were higher in IOLMaster 700 than Pentacam HR (p < 0.01). The width of the 95% limit of agreement was wide for all parameters. Conclusions: IOLMaster 700 showed better repeatability than Pentacam HR in chord mu, Px, Py, and pupil diameter values. Because there were statistically significant differences and a low level of agreement in chord mu and pupil diameter values between the two devices, they cannot be used interchangeably.

목차

Material and Methods
Results
Discussion
References

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

Na Hyun Kim,Hyun Jin Kim,Soo Chang Cho,Kyung Eun Han. (2023).Repeatability and Agreement of Chord Mu between Scheimpflug Tomography and Swept-Source Optical Coherence Tomography. The Korean Journal of Ophthalmology, 37 (6), 510-517

MLA

Na Hyun Kim,Hyun Jin Kim,Soo Chang Cho,Kyung Eun Han. "Repeatability and Agreement of Chord Mu between Scheimpflug Tomography and Swept-Source Optical Coherence Tomography." The Korean Journal of Ophthalmology, 37.6(2023): 510-517

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제