Introduction: The aim of this study was to demonstrate that the simulation surgery on rapid prototype (RP) model, which is based on the 3-dimensional computed tomography (3D CT) data taken before surgery, has the same accuracy as traditional orthograthic surgery with an intermediate splint, using an optoelectronic tracking navigation system.
Materials and Methods: Simulation surgery with the same treatment plan as the Le Fort I osteotomy on the patient was done on a RP model based on the 3D CT data of 12 patients who had undergone a Le Fort I osteotomy in the department of oral and maxillofacial surgery, Seoul National University Dental Hospital. The 12 distances between 4 points on the skull, such as both infraorbital foramen and both supraorbital foramen, and 3 points on maxilla, such as the contact point of both maxillary central incisors and mesiobuccal cuspal tip of both maxillary first molars, were tracked using an optoelectronic tracking navigation system. The distances before surgery were compared to evaluate the accuracy of the RP model and the distance changes of 3D CT image after surgery were compared with those of the RP model after simulation surgery.
Results: A paired t-test revealed a significant difference between the distances in the 3D CT image and RP model before surgery.(P<0.0001) On the other hand, Pearson s correlation coefficient, 0.995, revealed a significant positive correlation between the distances.(P<0.0001) There was a significant difference between the change in the distance of the 3D CT image and RP model in before and after surgery.(P<0.05) The Pearson s correlation coefficient was 0.13844, indicating positive correlation.(P<0.1) Conclusion: Theses results suggest that the simulation surgery of a Le Fort I osteotomy using an optoelectronic tracking navigation system is relatively accurate in comparing the pre-, and post-operative 3D CT data. Furthermore, the application of an optoelectronic tracking navigation system may be a predictable and efficient method in Le Fort I orthognathic surgery.
목차
Ⅰ. 서 론
Ⅱ. 연구 대상 및 방법
Ⅲ. 결 과
Ⅳ. 총괄 및 고찰
키워드
해당간행물 수록 논문
- 상악동측방접근법시 차폐막을 사용하지 않는 골창재위치술의 유용성 평가
- An expanded juvenile ossifying fibroma in maxillary sinus: a case report
- 하악 제3대구치 발치 시 합병증으로 발생할 수 있는 악하선과 설신경 손상: 증례보고
- 악골 내 거대 낭종성 병소의 감압술을 위해 고안된 장치를 이용한 치험례: 증례보고
- 광학추적항법장치를 이용한 르포씨 제1형 골절단 가상 수술의 정확성에 대한 연구
- 안면 비대칭환자의 하악골체부의 시상 단면에 관한 연구
- 자가치아골이식재를 이용한 골유도재생술: 증례보고
- 구강 편평세포암종에서의 암줄기세포 이론과 최신 지견
참고문헌
관련논문
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!