본문 바로가기

추천 검색어

실시간 인기 검색어

학술논문

광학추적항법장치를 이용한 르포씨 제1형 골절단 가상 수술의 정확성에 대한 연구

이용수 0

영문명
Accuracy of simulation surgery of Le Fort I osteotomy using optoelectronic tracking navigation system
발행기관
대한구강악안면외과학회
저자명
부연지(Yeon-Ji Bu)김성민(Soung-Min Kim) 김지연(Ji-Youn Kim) 박정민(Jung-Min Park) 명 훈(Hoon Myoung) 이종호(Jong-Ho Lee) 김명진(Myung-Jin Kim)
간행물 정보
『대한구강악안면외과학회지』대한구강악안면외과학회지 제37권 제2호, 114~121쪽, 전체 8쪽
주제분류
의약학 > 내과학
파일형태
PDF
발행일자
2011.04.30
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

Introduction: The aim of this study was to demonstrate that the simulation surgery on rapid prototype (RP) model, which is based on the 3-dimensional computed tomography (3D CT) data taken before surgery, has the same accuracy as traditional orthograthic surgery with an intermediate splint, using an optoelectronic tracking navigation system. Materials and Methods: Simulation surgery with the same treatment plan as the Le Fort I osteotomy on the patient was done on a RP model based on the 3D CT data of 12 patients who had undergone a Le Fort I osteotomy in the department of oral and maxillofacial surgery, Seoul National University Dental Hospital. The 12 distances between 4 points on the skull, such as both infraorbital foramen and both supraorbital foramen, and 3 points on maxilla, such as the contact point of both maxillary central incisors and mesiobuccal cuspal tip of both maxillary first molars, were tracked using an optoelectronic tracking navigation system. The distances before surgery were compared to evaluate the accuracy of the RP model and the distance changes of 3D CT image after surgery were compared with those of the RP model after simulation surgery. Results: A paired t-test revealed a significant difference between the distances in the 3D CT image and RP model before surgery.(P<0.0001) On the other hand, Pearson s correlation coefficient, 0.995, revealed a significant positive correlation between the distances.(P<0.0001) There was a significant difference between the change in the distance of the 3D CT image and RP model in before and after surgery.(P<0.05) The Pearson s correlation coefficient was 0.13844, indicating positive correlation.(P<0.1) Conclusion: Theses results suggest that the simulation surgery of a Le Fort I osteotomy using an optoelectronic tracking navigation system is relatively accurate in comparing the pre-, and post-operative 3D CT data. Furthermore, the application of an optoelectronic tracking navigation system may be a predictable and efficient method in Le Fort I orthognathic surgery.

목차

Ⅰ. 서 론
Ⅱ. 연구 대상 및 방법
Ⅲ. 결 과
Ⅳ. 총괄 및 고찰

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

부연지(Yeon-Ji Bu)김성민(Soung-Min Kim),김지연(Ji-Youn Kim),박정민(Jung-Min Park),명 훈(Hoon Myoung),이종호(Jong-Ho Lee),김명진(Myung-Jin Kim). (2011).광학추적항법장치를 이용한 르포씨 제1형 골절단 가상 수술의 정확성에 대한 연구. 대한구강악안면외과학회지, 37 (2), 114-121

MLA

부연지(Yeon-Ji Bu)김성민(Soung-Min Kim),김지연(Ji-Youn Kim),박정민(Jung-Min Park),명 훈(Hoon Myoung),이종호(Jong-Ho Lee),김명진(Myung-Jin Kim). "광학추적항법장치를 이용한 르포씨 제1형 골절단 가상 수술의 정확성에 대한 연구." 대한구강악안면외과학회지, 37.2(2011): 114-121

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제